Knowledgebase

Lafayette Instrument Logo

Knowledgebase

Why does my EDA component sometimes display a flat line?

Question Last Updated on: 6/7/2007

Just as on an analog instrument the EDA will sometimes display a flat line during testing. It is always recommended to make sure that your subject has washed their hands with warm water and soap before the test and make sure they have not placed makeup or lotion on their hands. Try not to use soaps or prep pads with alcohol. If you have already recorded a chart you can also refer to the EDA Ohms table on the right side of the chart. If there is almost no difference between Average, Min. and Max. resistance or the resistance reaches 2.0 M (megaohms) then the subject is either not reacting or is reacting beyond the limits of what the DAS can measure. If your problem persists please give us a call.



Search Knowledgebase

Submit Question

Unable find an answer to the question that you looking for in our FAQ section? Submit your question directly to our customer service representatives.

             Join our email list for up-to-date product news             Sign Me Up!

Copyright © 2009 - 2024 Lafayette Instrument Company. All Rights Reserved.